Technical note

A Practical 5-Step Guide to Verifying WIKA A-10 Pressure Transmitter Accuracy Before Installation

Who Should Use This Checklist (And When)

This is for engineers and technicians staring at a WIKA A-10 pressure transmitter, wondering if the stated accuracy on the box will hold up in their actual process. It's for the moment right before you install—when a five-minute check can save a five-hour problem.

I've been handling instrumentation orders for 6 years now. In my first year (2017), I made the classic mistake of trusting the datasheet headline number without reading the fine print. The result? A $3,200 order had to be re-spec'd because the transmitter's accuracy at the extreme of its range didn't meet our process requirements. This checklist is what I wish I'd had that day.

Three things you'll get from this:

  • A step-by-step process to cross-reference published specs with your actual conditions.
  • The one parameter most people skip—and why it costs you.
  • How to use WIKA's technical documentation to avoid a costly mismatch.

Step 1: Deconstruct the Datasheet Accuracy Statement

Every WIKA A-10 datasheet has an accuracy spec. It looks simple. It's not.

Let's take a typical line: "Accuracy: ≤ 0.5 % of span." Sounds straightforward, right? But there are two traps you need to check.

Trap 1: Is it % of span or % of reading? The A-10 usually specifies % of span. That means the error is a fixed percentage of the transmitter's full range, not the measured value. If you're measuring at the low end of the range, the error is larger relative to your reading. I once saw a spec claim "0.5% accuracy"—but it was 0.5% of span, and our process operated at 10% of span. The effective accuracy was 5% of reading. Acceptable? Not for our control loop.

Trap 2: What conditions are assumed? The headline number is valid only under reference conditions (usually 23°C ±2°C, no vibration, specific supply voltage). Your installation will not be those conditions.

Your task here:

  • Find the accuracy statement on the datasheet.
  • Note whether it's % of span, % of reading, or % of URL (upper range limit).
  • Look for the reference conditions footnote. Write them down.

This took me about 45 seconds once I learned where to look. The datasheet for the A-10 has a clear table on the second page.

Step 2: Calculate Your Worst-Case Error (At Your Operating Point)

This is where the math happens. Don't skip it.

Formula:

Worst-case error = (Accuracy % / 100) × (Full range)

Example:

  • Transmitter range: 0-100 psi
  • Accuracy: 0.5% of span
  • Your operating point: 20 psi
  • Worst-case error = (0.5 / 100) × 100 psi = 0.5 psi
  • That 0.5 psi is 2.5% of your actual reading at 20 psi.

Does that matter for your process? Maybe. For a tank level application, 2.5% might be fine. For a critical pressure safety interlock? Probably not.

Pro tip: I keep a quick calculation in my notebook. It's faster than pulling up Excel every time. Just a simple table with common operating points and their effective error.

Step 3: Apply Temperature Effects (The One Most People Skip)

Here's the step I ignored in 2017. The A-10 datasheet will have a temperature coefficient—usually something like "≤ ±0.2% of span per 10K change in ambient temperature."

This is where your installation environment matters. If your process is in a conditioned control room, fine. If it's on a sunny pipeline in Houston in August, you have a problem.

How to calculate temperature error:

  1. Find your expected temperature variation from reference conditions. Reference is usually 23°C. If your process area hits 40°C, that's +17K.
  2. Divide by the spec interval (10K). That gives you ~1.7 intervals.
  3. Multiply by the temperature coefficient. 1.7 × 0.2% = 0.34% additional error.
  4. Add that to your basic accuracy. 0.5% + 0.34% = 0.84% total potential error.

Suddenly, your "0.5%" transmitter is effectively 0.84% in your installation. And that's before we talk about vibration or supply voltage.

I still kick myself for not doing this on that $3,200 order. If I'd applied temperature effects first, I'd have ordered the A-10 with the extended temperature compensation option. $45 more on the PO. Worth every penny.

Step 4: Check the Calibration Certificate Against Your Application

Every WIKA A-10 comes with either a standard factory calibration or, if you paid for it, a traceable certificate. Most people just file it. Don't.

Here's what to verify:

  • Is the calibration performed at the range you ordered? A 0-100 psi transmitter calibrated at 0-100 psi is fine. One that was calibrated as a standard 0-100 psi but your application requires a 0-50 psi turndown? The accuracy spec changes. Check the datasheet for turndown effects.
  • Are the test points relevant? The standard certificate usually does 5 points (0%, 25%, 50%, 75%, 100%). If your operating point is at 10% of range, you're outside the verified range. Not ideal. Consider requesting additional points.
  • Is the as-found data acceptable? Compare the as-found calibration results against the claimed accuracy. If the transmitter passed but barely, and your process is tight, maybe this unit isn't the best one.

I have mixed feelings about this step. On one hand, it's a 5-minute check. On the other, if you don't have the certificate handy (it's often emailed separately from the physical shipment), it's easy to skip. I make it a rule now: the transmitter doesn't leave the receiving dock until I've seen the cert.

Step 5: Perform a Quick Incoming Inspection (The Confidence Check)

This is your last line of defense. You don't need a full re-calibration—just a functional check to catch gross errors.

What you'll need:

  • A stable pressure source (a hand pump or deadweight tester).
  • A reference standard (at least 4x more accurate than the transmitter).
  • 24V DC power supply (or whatever the A-10's loop-powered voltage requires).

The 3-point check:

  1. Apply 0% pressure. Read the output. Should be 4 mA (or 0 VDC). Note the deviation.
  2. Apply 50% pressure. Read the output. Should be 12 mA (or 50% of voltage span).
  3. Apply 100% pressure. Read the output. Should be 20 mA (or full voltage span).

Each point should be within the accuracy spec we calculated back in Step 1 and Step 3 combined. If it's not? That's your flag. Don't install it—return it.

One more thing: Check the physical condition. Bent process connection? Scratched diaphragm? It might have been damaged in shipping. I caught one once where the pressure port was dented. The inside looked fine, but I sent it back anyway. Not worth the risk.

Common Mistakes to Watch For

I've made most of these. You don't have to.

  • Trusting the headline number. Accuracy is never just one number. You need the full context.
  • Skipping the temperature calculation. This is the one I see most engineers miss. Your installation temperature matters.
  • Ignoring the calibration certificate. It's not just paperwork. It's your first piece of evidence that the transmitter will do what it says.
  • Not documenting your checks. I use a simple form now. Date, transmitter serial number, three test points, pass/fail. It's saved us in audits more than once.

Quick final thought: The WIKA A-10 is a solid transmitter. I've installed dozens. But solid doesn't mean bulletproof. A five-minute check—using this list—can prevent a day of troubleshooting and a potential process upset. That's worth the time.

Jane Smith

Jane Smith

I’m Jane Smith, a senior content writer with over 15 years of experience in the packaging and printing industry. I specialize in writing about the latest trends, technologies, and best practices in packaging design, sustainability, and printing techniques. My goal is to help businesses understand complex printing processes and design solutions that enhance both product packaging and brand visibility.